Device Priming - Priming Set With a Pressure Sensing Disc (continued)

8. Install pressure sensing disc.

9. Press EXIT soft key to return to previous programming screen.

10. Complete programming selected infusion type and unclamp set.

11. Start infusion and then attach administration set to patient.

Note: Volume used for priming is displayed on priming screen. This volume is not added to VTBI or VI. Starting the infusion before attaching set to patient will minimize any potential bolus that could be released from pressure built up in the set due to normal priming and syringe loading.