Alaris® Syringe module
Topic 3: Working with an Infusion


Page 2 of 10

Glossary

Resources

Back

Next

Menu

Exit


ALL Mode

When ALL Mode is established in a profile, the entire contents of the syringe will be delivered. If the infusion you are programming does not require that the entire contents be delivered, you can elect to clear the ALL Mode and enter a VTBI.

  1. Press the CHANNEL SELECT key on the Alaris® Syringe module.
  2. On the setup screen, press the VTBI soft key.
  3. Enter a VTBI using the numeric keys.