Alaris® Syringe module
Topic 2: Preparing an Infusion


Page 11 of 21

Glossary

Resources

Back

Next

Menu

Exit


Device Priming - Priming Set Without a Pressure Sensing Disc (continued)

5. Press and hold PRIME soft key.

6. When priming is complete, release PRIME soft key.

7. Press EXIT soft key to return to previous programming screen.

8. Complete programming selected infusion type and unclamp set.

9. Start infusion and then attach administration set to patient.

Note: Volume used for priming is displayed on priming screen. This volume is not added to VTBI or VI. Starting the infusion before attaching set to patient will minimize any potential bolus that could be released from pressure built up in the set due to normal priming and syringe loading.