Selecting a Pressure Limit with Pressure Sensing Disc Installed

4. Press the Auto Pressure soft key or the Change Value soft key.

5. If you selected Change Value, use the numeric keys to enter a specific mmHg value in increments as small as 1 mmHg.

6. Verify your selection and press the CONFIRM soft key.

Note: Auto Pressure automatically sets the limit as follows:

  • If the current pressure is 100 mmHg or less, the system adds 30 mmHg to the current pressure.
  • If the current pressure is greater than 100 mmHg, the system adds 30% to the current pressure.