Alaris® Pump module
Topic 3: Primary and Secondary Infusions


Page 10 of 18

Glossary

Resources

Back

Next

Menu

Exit


Programming a Secondary Infusion with Guardrails® Suite MX Protection

5. Verify parameters are correct and press the NEXT soft key to confirm.

6. Use the numeric keys to enter the secondary infusion VTBI. (Note: The screen defaults to a Volume/Duration infusion, but you can change it to a Rate/Volume infusion by pressing the RATE/VOLUME soft key.)

7. Based on your hospital's guidelines, an initial starting duration may be pre-populated. To change the starting duration, press theDURATION active soft key and enter the desired amount using the numeric keys.

8. Verify the parameters are correct, open the clamp on the secondary administration set, and press the START soft key.