Secondary Infusion Status

During a secondary infusion:

  • The Green Infusing Status Indicator illuminates.
  • The secondary rate appears in the Channel Rate Display.
  • The secondary VTBI counts down on the Main Display.
  • SECONDARY and the drug name and dose scroll in the Channel Message Display.

At the completion of a secondary infusion:

  • The switchover alert sounds with six beeps (unless disabled in the system configuration).
  • The primary infusion resumes.
  • The primary rate displays.