Helpful Hints
- Always swab the top of the SmartSite® needle-free valve
port with a sterile 70% isopropyl alcohol wipe (1-2
seconds) and allow to dry (approximately 30 seconds).
- Replace set every 72 hours or 100 activations, whichever occurs
first. For infusions of blood, blood products, or lipid emulsions,
replace
every 24 hours.
- Prime the SmartSite® needle-free valve
prior to connecting it to the patient's catheter or extension set.
- Some medications may continue to be supplied in pre-filled syringes
with non-removable needles. Please follow the recommendations
below only when using a syringe with a non-removable needle.
- The SmartSite® needle-free valve will leak if punctured with a
needle. In an emergency, a standard PRN adapter can be added to the
SmartSite® needle-free valve to allow access with a needle.
- Ensure that when accessing the PRN adapter with a needle that the
needle does not come in contact with the SmartSite® needle-free valve.
- As a temporary measure for a punctured SmartSite® needle-free valve, you
can add an additional SmartSite® needle-free valve onto the damaged/punctured
SmartSite® needle-free valve. However, once the integrity of the SmartSite® needle-free valve has been compromised, it needs to be replaced.
|
|