Manual Priming With a Pressure Sensing
Disc
To manually prime an administration set with a pressure sensing disc:
- Ensure administration set is not connected to patient and that
device is as close to level of patient as possible (patient should
be in line with CHANNEL SELECT key on Alaris® Syringe
module).
- Hold pressure sensing disc between two fingers and
invert disc so patient side is up.
- Slowly prime set while gently massaging disc to ensure all air
is expelled. Note that disc must remain inverted only until
all air
is expelled. Continue to gently massage disc to ensure
it is not over- or under-filled.
- When priming is complete and no air
exists, clamp set.
- Load syringe and pressure sensing disc.
- Complete programming selected infusion and unclamp set.
- Start infusion and then attach set to patient.
Note: Starting the infusion before attaching
set to patient will minimize any potential bolus that could be released
from pressure built up in the set due to normal priming and syringe
loading.
|